Intended Use
Product Introduction
General Description
Malaria is caused by a protozoan which invades human red blood cells.1Malaria is one of the world's most prevalent diseases. According to the WHO, the worldwide prevalence of the disease is estimated to be300-500 million cases and over 1 million deaths each year. Most of these victims are infants, young children. Over half of the world's population lives in malarious areas. Microscopic analysis of appropriately stained thick and thin blood smears has been the standard diagnostic technique for identifying malaria infections for more than a century.2The technique is capable of accurate and reliable diagnosis when performed by skilled microscopists using defined protocols. The skill of the microscopist and use of proven and defined procedures, frequently present the greatest obstacles to fully achieving the potential accuracy of microscopic diagnosis. Although there is a logistical burden associated with performing a time-intensive, labor-intensive, and equipment-intensive procedure such as diagnostic microscopy, it is the training required to establish and sustain competent performance of microscopy that poses the greatest difficulty in employing this diagnostic technology. The Malaria P.f. Rapid Test Device (Whole Blood) is a rapid test to qualitatively detect the presence of the P.f. antigen.
Principle
The Malaria P.f. Rapid Test Device (Whole Blood) is a qualitative, membrane based immunoassay for the detection of P.f. antigen in whole blood. The membrane is precoated with P.f. antibody. During testing, the whole blood specimen reacts with the dye conjugate, which has been pre-coated in the Test Device. The mixture then migrates upward on the membrane chromatographically by capillary action and reacts with P.f. antibody on the membrane on the test line. If the specimen contains P.f. antigen, a colored line will appear in the test region. The absence of the colored line in test region indicates that the specimen does not contain P.f. antigen. To serve as a procedure control, a colored line will always appear in the control region indicating that proper volume of specimen has been added and membrane wicking has occurred.
Reagents And Materials Provided
The Test Device contains monoclonal anti-P.falciparumantibodies coated on the membrane
• Test Devices • Disposable specimen droppers
• Buffer • Package insert
Materials Required But Not Supplied
• Pipette and disposable tips(optional) • Specimen collection containersSpecimen Collection And Preparation
• The Malaria P.f. Rapid Test Device (Whole Blood) can be performed using whole blood.Reconstitution And Storage
The kit can be stored at room temperature or refrigerated (2-30°C). The Test Device is stable through the expiration date printed on the sealed pouch. The Test Device must remain in the sealed pouch until use. DO NOT FREEZE. Do not use beyond the expiration date.
Assay Procedure
Interpretation of Results
Quality Control
Internal procedural controls are included in the test. A colored line appearing in the control region (C) is an internal positive procedural control. It confirms sufficient specimen volume and correct procedural technique. Control standards are not supplied with this kit; however, it is recommended that positive and negative controls be tested as a good laboratory practice to confirm the test procedure and to verify proper test performance.
Reference Values
The Malaria P.f. Rapid Test Device (Whole Blood) has been compared with traditional thick or thin blood smears microscopic analysis. The correlation between the two systems is 99.7%.
Performance Characteristics
##Sensitivity##
The Malaria P.f. Rapid Test Device (Whole Blood) has been tested with thin or thick blood smears on clinical specimens. The results show that the sensitivity of the Malaria P.f. Rapid Test Device (Whole Blood) is >99.0% relative to blood smears.
##Specificity##
The Malaria P.f. Rapid Test Device (Whole Blood) uses an antibody that is highly specific for Malaria P.f. antigen in whole blood. The results show that the specificity of the Malaria P.f. Rapid Test Device (Whole Blood) is 99.7% relative to blood smears.
Relative Sensitivity: >99.0% (91.8%-100.0%)* Relative Specificity: 99.7% (98.3%-100.0%)* Accuracy: 99.7% (98.5%-100.0%)* * 95% Confidence Interval
##Precision##
Intra Assay
Within-run precision has been determined by using 10 replicates of twelve specimens containing negative, low positive and high positive samples. The negative and positive values were correctly identified >99% of the time.
Inter Assay
Between-run precision has been determined by using the same twelve specimens of negative, low positive and high positive of 10 independent assays and with three different lots of the Malaria P.f. Rapid Test Device (Whole Blood). The negative and positive values were correctly identified >99% of the time.
Precautions